Klein-tools DTL304 User Manual Page 7

  • Download
  • Add to my manuals
  • Print
  • Page
    / 36
  • Table of contents
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 6
7
Press “
” or
until the display shows “
.
Press “
ENTER
” to display “
” or “
” in the time display.
If “
” is displayed, press “
” or
to select “
”.
Press “
ENTER
” to display choice number 3 in the secondary
display and
00:30
h:m
in the time display.
Press “
” or
until the display shows the Auto-Power-
Off Time choice number you want, and then press “
ENTER
” to
store the Auto-Power-Off setting in memory.
Auto-Power-Off mode is automatically disabled in MIN/MAX
and logging modes.
H. Change Logging Interval
The logging interval determines how often the thermometer
stores logged readings in memory. You choose the length of
the logging interval. The thermometer stores logged readings
at the end of each logging interval.
Select logging intervals
Press “
” or
until the display shows
and “
LOG
.
Press “
ENTER
” to display the logging interval choices.
Press “
” or
until the display shows the logging
interval you want, and then press “
ENTER
” to select.
For user-defined logging interval (1 sec. to 23 hrs : 59 mins).
Press “
” or
until the display shows “
h:m
” or “
m:s
”,
and then press “
ENTER
” to select. The last number in the
time display blinks.
Press “
” or
until the logging interval you want
appears on the display, and then press “
ENTER
” to select.
Holding down “
” or
causes the number to change
more quickly.
I. Setting Temperature Offset
You can adjust the thermometer’s readings to compensate
for the errors of a specific thermocouple. The allowable
compensation range is ±5.0°C or K, and ±9.0°F.
The DTL304
allows you to store individual offsets for T1, T2, T3 and T4.
Press “
” or
until the display shows “
OFFSET
”.
Press “
ENTER
” to display T1 reading (blinking), T1 offset.
Press “
” or
to display
T1, T2, T3 and T4
in the primary
display to select 1 input.
Press “
” or
until the primary display shows the
correct reading, and then press “
ENTER
” to store the offset
setting in memory.
NOTE: Reset the offset to 0.0 when it is no longer needed.
NOTE: If T1, T2, T3 or T4 has no offset, “ OFFSET ”will not be
displayed.
J. Resetting The Thermometer
To restore to factory settings, press “
” or
until
the display shows “
” then press “
ENTER
to display
“ in the primary display. Press “
EXIT
” to exit setup.
FUNCTION INSTRUCTIONS
1. Connecting Thermocouples
Thermocouples are color coded based on the ANSI color code:
Plug a thermocouple into the input terminal(s).
Set the thermometer for the correct thermocouple type. To
change the thermocouple type, see:
Set Up Instructions
section,
Set Up Instructions section, Set Up Instructions
F: Thermocouple Type
1.
:01
sec
2.
:10
sec
3.
1:00
min
4.
10:00
min
User Defined
J Type =
Black
K Type =
Yellow
T Type =
Blue
E Type =
Purple
Page view 6