Klein-tools DTL302 User Manual Page 10

  • Download
  • Add to my manuals
  • Print
  • Page
    / 36
  • Table of contents
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 9
10
ENGLISH
ELECTRICAL SPECIFICATIONS
Temperature
Range
J-type
-346°F~2192°F
-210°C~1200°C
64K ~ 1472K
K-Type
-328°F~2498°F
-200°C~1370°C
74K ~ 1642K
T-type -328°F~752°F -200°C~400°C 74K ~ 673K
E-Type -238°F~1832°F -150°C~1000°C 124K ~ 1273K
Display
Resolution
J-type
0.1°F<1000°F
0.1°C<550°C
0.1K<824K
1°F
1000°F
1°C
550°C
1K
824K
K-Type
0.1°F<1000°F
0.1°C<750°C
0.1K<1000K
1°F
1000°F
1°C
750°C
1K
1000K
T-type
0.1°F<752°F
0.1°C<400°C
0.1K<673K
N/A
N/A
N/A
E-Type
0.1°F<750°F
0.1°C<400°C
0.1K<673K
1°F
750°F
1°C
400°C
1K
673K
Temperature
Accuracy*
J, K, E
Type
±[0.1% + 1.4˚F(0.8˚C)]
<-148˚F(-100˚C) add 0.2% of reading
T Type
±[0.3% + 2.7˚F(1.5˚C)]
<-148˚F(- 100˚C) add 0.5% of reading
Temperature
Coeffi cient
0.01% of reading 0.1°F per °F(+0.05°C per °C) for
<+64°F (+18°C) or +82°F (+28°C)
[Below -148°F (-100°C): add 0.05% of reading for J,
K, and E-type; and 0.1% of reading for T-type]
Real Time
Clock Tolerance
About 1 second per day
Max
Differential
Common Mode
Voltage
1V Maximum voltage difference
(between any pair of inputs)
Temperature
Scale
ITS-90 (International Temperature Scale of 1990)
Standards
N.I.S.T. Monograph 175 revised to ITS-90
* Accuracy is specified for ambient temperatures between 64°F and
82°F (18°C and 28°C) for a period of 1 year. The above specifications
do not include thermocouple error.